王惠, 李国宏, 蔡蕾, 崔雪华, 任震晴. 自制充气垫在甲状腺手术中的应用研究[J]. 实用临床医药杂志, 2014, (22): 55-57,67. DOI: 10.7619/jcmp.201422017
引用本文: 王惠, 李国宏, 蔡蕾, 崔雪华, 任震晴. 自制充气垫在甲状腺手术中的应用研究[J]. 实用临床医药杂志, 2014, (22): 55-57,67. DOI: 10.7619/jcmp.201422017
WANG Hui, LI Guohong, CAI Lei, CUI Xuehua, REN Zhenqing. Application of self-made inflatable mattress in thyroid surgery[J]. Journal of Clinical Medicine in Practice, 2014, (22): 55-57,67. DOI: 10.7619/jcmp.201422017
Citation: WANG Hui, LI Guohong, CAI Lei, CUI Xuehua, REN Zhenqing. Application of self-made inflatable mattress in thyroid surgery[J]. Journal of Clinical Medicine in Practice, 2014, (22): 55-57,67. DOI: 10.7619/jcmp.201422017

自制充气垫在甲状腺手术中的应用研究

Application of self-made inflatable mattress in thyroid surgery

  • 摘要: 目的:探讨自制充气肩垫在甲状腺手术中的应用效果。方法将100例甲状腺手术患者随机分为对照组和试验组,各50例。对照组采用(50×43×8)cm~(50×43×12)cm、斜坡成20°的海绵体位垫进行手术,试验组采用自制同尺寸可控充气体位垫进行手术,比较2组手术患者体位摆放前、摆放时、恢复平卧位后的血压及心率变化,患者的舒适度及医生操作满意度等指标。结果对照组和试验组在手术体位摆放前血压、心率无显著差异(P >0.05),手术体位摆放后试验组测量血压、心率低于对照组,差异有统计学意义(P <0.05);手术患者的舒适度及医生操作满意试验组高于对照组,差异有统计学意义(P <0.05)。结论应用自制可控充气垫能避免摆放体位搬动时生命体征的变化,保持血流动力学的稳定,提高手术患者的舒适度及医生操作满意度。

     

    Abstract: ABSTRACT:Objective To explore the application effect of self-made inflatable mattress in thyroid surgery.Methods A total of 100 patients undergoing thyroid surgeries were randomly di-vided into control group and test groups with 50 cases in each of group.The control group was per-formed surgery with foam position pad on mattress with the area from (50×43×8)cm to (50×43 ×12)cm at 20 degree angles.The test group was performed surgery with controllable self-made inflatable mattress.Blood pressure,heart rate in different positions,comfort degree of surgery pa-tient and satisfaction index was recorded.Results There was no significant difference (P >0.05) between control group and test group in blood pressure and heart rate.Blood pressure and heart rate of patients of surgery in test group were lower than that of control group (P <0.05).There were statistically significant differences since comfort degree of surgery patient and satisfaction in-dex in control group were less than that of test group (P <0.05).Conclusion Controllable self-made inflatable mattress could avoid the changes in vital signs and remain a stable hemodynamics and improve comfort degree of patients and satisfaction index.

     

/