ZHANG Kailian, ZHANG Fen, LIN Fei, LI Ya, ZHAO Biao, YANG Yi. Preliminary analysis in body position error of intensity modulated radiotherapy after breast conserving surgery underdouble-tagging thermoplastic body film fixation[J]. Journal of Clinical Medicine in Practice, 2015, (17): 51-53. DOI: 10.7619/jcmp.201517016
Citation: ZHANG Kailian, ZHANG Fen, LIN Fei, LI Ya, ZHAO Biao, YANG Yi. Preliminary analysis in body position error of intensity modulated radiotherapy after breast conserving surgery underdouble-tagging thermoplastic body film fixation[J]. Journal of Clinical Medicine in Practice, 2015, (17): 51-53. DOI: 10.7619/jcmp.201517016

热塑体膜固定双重标记下保乳术后调强放疗移位误差的初步分析

Preliminary analysis in body position error of intensity modulated radiotherapy after breast conserving surgery underdouble-tagging thermoplastic body film fixation

  • 摘要: 目的 探讨热塑体膜固定双重标记在乳腺癌保乳术后调强放疗中应用的周期性误差.方法 对20例乳腺癌保乳术后热塑体膜固定双重标记调强放疗患者分别在治疗开始后l、2、3周采用模拟定位机测量患者的左右、头足、前后方向和乳房的重复摆位误差.结果 第1周误差较大,而第2、3周的误差相对较小,第1、2、3周相比较差异有统计学意义(P<0.05);各个方向的摆位和银夹移位误差均≤2.5 mm.结论 热塑体膜固定双重标记技术可减少摆位误差,在乳腺癌保乳术后调强放疗中有较好稳定性和重复性.

     

/