李文超, 宋永红, 陈剑锋. PK7300全自动血型仪在血型检测中的应用分析[J]. 实用临床医药杂志, 2021, 25(9): 32-36. DOI: 10.7619/jcmp.20210170
引用本文: 李文超, 宋永红, 陈剑锋. PK7300全自动血型仪在血型检测中的应用分析[J]. 实用临床医药杂志, 2021, 25(9): 32-36. DOI: 10.7619/jcmp.20210170
LI Wenchao, SONG Yonghong, CHEN Jianfeng. Application of PK7300 automatic blood grouping analyzer in blood grouping test[J]. Journal of Clinical Medicine in Practice, 2021, 25(9): 32-36. DOI: 10.7619/jcmp.20210170
Citation: LI Wenchao, SONG Yonghong, CHEN Jianfeng. Application of PK7300 automatic blood grouping analyzer in blood grouping test[J]. Journal of Clinical Medicine in Practice, 2021, 25(9): 32-36. DOI: 10.7619/jcmp.20210170

PK7300全自动血型仪在血型检测中的应用分析

Application of PK7300 automatic blood grouping analyzer in blood grouping test

  • 摘要:
      目的  分析PK7300全自动血型仪在血型检测中的应用价值。
      方法  使用奥林巴斯PK7300全自动血型仪检测献血者血液样本ABO及RhD血型,对判定失败样本进行手工复核,将异常血型及初筛Rh阴性血型样本送血型参比室鉴定。
      结果  2016—2019年共检测402 649份样本,血型分析仪一次性判读成功399 841份(99.30%),判读失败2 808份(0.70%);送检RhD血型鉴定样本2 730份,确认阴性2 644份(96.85%),阳性86份(3.15%);送检ABO正反定型不符样本101份,鉴定显示血型正常44份,AB亚型26份,含MNSs-抗-M抗体31份;送检单独O细胞凝集反应样本24份,鉴定显示血型正常7份,含MNSs-抗-M抗体14份,冷自身抗体2份,温自身抗体1份。
      结论  PK7300血型仪能够检出一定数量的亚型和含MNSs-抗-M抗体的样本,具有操作简单、自动化程度高、准确率高和使用试剂量少等特点,可提升血型检测的工作效率和检测质量,适用于大批量血型检测。

     

    Abstract:
      Objective  To analyze the application value of PK7300 automatic blood grouping analyzer in blood grouping identification.
      Methods  Blood specimens from blood donors were analyzed by Olympus PK7300 blood grouping analyzer to identify the ABO and RhD blood groups. Samples with failed interpretation were retested by manual method. Abnormal blood types or initially screened as Rh-negative specimens were sent to the Blood Group Reference Laboratory of our center for further identification.
      Results  A total of 402 649 blood specimens during 2016 to 2019 were analyzed. In sum, there were 399 841 (99.30%) samples successfully identified at the first time, and 2 808 (0.70%) failed. A total of 2 730 RhD samples were sent for further identification, including 2 644 (96.85%) Rh-negative samples, and 86 (3.15%) Rh-positive ones. There were 101 samples with ABO discrepancies, among which 44 were identified as normal ABO blood group, 26 as AB subtypes, and 31 contained MNSs-anti-M antibodies. Out of 24 samples of O-cell agglutination, 7 were identified as normal blood group, 14 contained MNSs-anti-M antibodies, 2 contained cold autoantibodies, and 1 contained warm autoantibodies.
      Conclusion  PK7300 automatic blood grouping analyzer can determine a certain number of ABO subtypes and MNSs-anti-M antibodies, and has the features of simple operation, high automation, high accuracy, and less reagent consuming, which can improve the efficiency and quality of blood group identification, and is suitable for large-scale blood group screening.

     

/